Talk to an engineer

Research · in vivo tested

WPTxHeart: wireless power for a left ventricular assist device

A subcutaneous wireless power system that replaces the percutaneous driveline of an LVAD, tested in vivo on an animal model with the cardiac surgery unit of the San Camillo-Forlanini hospital in Rome: one of the first experiments of this kind in Italy.

Device
Left ventricular assist device (LVAD)
Configuration
External TX coil on the skin, subcutaneous RX coil in the abdomen
Implanted unit
RX coil with integrated battery pack and control electronics
Maturity
TRL 6 prototype, in vivo test on an animal model
Clinical partner
Cardiac surgery team of Prof. Francesco Musumeci and Dr. Andrea Montalto, San Camillo-Forlanini hospital, Rome
Power class
Low
WPTxHeart: wireless power for a left ventricular assist device: WPTxHeart pilot

01Context

An LVAD is an electric pump attached to the heart that helps the left ventricle push blood to the rest of the body. Today it bridges end-stage heart failure patients to a transplant; with too few organs available, it is becoming a long-term destination therapy.

02Challenge

The pump is powered by external batteries through a percutaneous driveline. The exit site of that cable is the entry point for infections: about 80% of LVAD infections are caused by this physical connection. There is no internal back-up battery, the cable weighs on the patient’s daily life, and pump delays or missed restarts remain a risk.

03Solution

  1. Power through the skin

    A transmitting coil worn on the abdomen powers a receiving coil implanted under the skin, replacing the section of the driveline that crosses the skin.

  2. Fewer implanted parts

    Internal battery and electronics sit inside the receiving coil, with an emergency connector in the subcutaneous unit.

  3. Safety first

    Electromagnetic exposure assessed on anatomical human models against ICNIRP limits, with temperature rise monitored on coil and electronics.

  4. Remote telemetry

    The external unit streams system data over a mobile link to a remote dashboard.

04Results

  • In vivo test on an animal model: one of the first experiments of this kind in Italy.
  • Normal operation of the pump with and without the transmitting coil in place.
  • Temperature rise of coil and electronics well below the limits.
  • Clinical parameters identical to those of the pump running on its driveline.
  • Next step: smaller components for a fully implantable system.

How it worksToday vs WPTxHeart

No cable through the skin.

TODAYPower through a percutaneous driveline OUTSIDESKININSIDE CONTROL INFECTION RISK~80% start at the exit site Driveline LVAD pump WPTxHEARTNo cable through the skin OUTSIDESKININSIDE CONTROL TX coilon the skin INTACT SKIN RX coil+ battery Internal cableto the pump LVAD pump

A coil on the skin powers a receiver with its battery under the skin; an internal cable takes the power to the pump.

Better for the patient.

Infections

No exit site, no way in for infection

About 80% of LVAD infections start where the driveline exits the skin. Here power crosses intact skin, so that entry point is gone.

Internal battery

Keeps running without the transmitter

A battery inside the implant keeps the pump running for a while, even with the external coil removed.

Quality of life

No cable tied to the body

No cable out of the abdomen and no wound to dress every day, for patients who live with the device for years.

~80%
of LVAD infections start at the driveline
In vivo
animal-model test, among the first of its kind in Italy
=
clinical parameters, same as with the driveline

On site4 photos

With Fondazione San Camillo-Forlanini

Next projectPacemaker researchWireless charging for pacemakersRead the case

Similar challenge?

Tell us about your vehicles, robots or site. An engineer will get back to you.

Talk to an engineer