Infections
No exit site, no way in for infectionAbout 80% of LVAD infections start where the driveline exits the skin. Here power crosses intact skin, so that entry point is gone.
Research · in vivo tested
A subcutaneous wireless power system that replaces the percutaneous driveline of an LVAD, tested in vivo on an animal model with the cardiac surgery unit of the San Camillo-Forlanini hospital in Rome: one of the first experiments of this kind in Italy.

01Context
An LVAD is an electric pump attached to the heart that helps the left ventricle push blood to the rest of the body. Today it bridges end-stage heart failure patients to a transplant; with too few organs available, it is becoming a long-term destination therapy.
02Challenge
The pump is powered by external batteries through a percutaneous driveline. The exit site of that cable is the entry point for infections: about 80% of LVAD infections are caused by this physical connection. There is no internal back-up battery, the cable weighs on the patient’s daily life, and pump delays or missed restarts remain a risk.
03Solution
A transmitting coil worn on the abdomen powers a receiving coil implanted under the skin, replacing the section of the driveline that crosses the skin.
Internal battery and electronics sit inside the receiving coil, with an emergency connector in the subcutaneous unit.
Electromagnetic exposure assessed on anatomical human models against ICNIRP limits, with temperature rise monitored on coil and electronics.
The external unit streams system data over a mobile link to a remote dashboard.
04Results
A coil on the skin powers a receiver with its battery under the skin; an internal cable takes the power to the pump.
Infections
No exit site, no way in for infectionAbout 80% of LVAD infections start where the driveline exits the skin. Here power crosses intact skin, so that entry point is gone.
Internal battery
Keeps running without the transmitterA battery inside the implant keeps the pump running for a while, even with the external coil removed.
Quality of life
No cable tied to the bodyNo cable out of the abdomen and no wound to dress every day, for patients who live with the device for years.
Test bench: the pump runs on the WPT system, with no percutaneous driveline
A conventional LVAD today: pump, batteries, external controller and the driveline through the skin
In vivo test with the cardiac surgery team of the San Camillo-Forlanini hospital, Rome
Prof. Francesco Musumeci and Dr. Andrea Montalto’s team during the implant
With Fondazione San Camillo-Forlanini
Next projectPacemaker researchWireless charging for pacemakersRead the case
Similar challenge?
Tell us about your vehicles, robots or site. An engineer will get back to you.